Single Use Medical Device Reprocessing

Single-use medical device reprocessing is the disinfection, cleaning, remanufacturing, testing, packaging and labeling, and sterilization among other steps, of a used, (or, in some cases, a device opened from its original packaging but unused), medical device to be put in service again. All reprocessed medical devices originally labeled for single use in the United States are subject to U.S. Food and Drug Administration (FDA) manufacturing requirements and must meet strict cleaning, functionality, and sterility specifications prior to use.[1] Although first regulated in the U.S., the reprocessing of medical devices, particularly those that are labeled “Single Use Device” (SUDs), is a global practice with countries in Europe, Asia, Africa, and North America actively engaged in reprocessing.[2] Currently, approximately 2% of all SUDs on the U.S. market are eligible for reprocessing by a qualified third-party vendor.[3] The U.S. revenue for reprocessed devices (not SUDs) is estimated to be around $400 million annually.[4]

History of reprocessing in the United States

The practice of reusing medical devices labeled for only one use began in hospitals in the late 1970s.[5] After a thorough review by the U.S. FDA in 1999 and 2000,[5] the agency released a guidance document for reprocessed SUDs that began regulating the sale of these reprocessed devices on the market,[6] under the condition that third-party reprocessors would be treated as the manufacturer and would meet the same criteria as the original equipment manufacturers (OEMs) of the medical device.[6] Following the implementation of FDA regulation, the U.S. Congress codified these standards and other requirements in the Medical Device User Fee Act of 2002.[7]

The Single Use Label

Original equipment manufacturers undergo a thorough analysis to determine whether a device should be single-use or reusable. Factors in determining to label a device single-use include but are not limited to possible infection/sterility concerns and the risk of reduced efficacy. Depending on the device and its design, failure to completely clean a single-use device could lead to cross patient infection or there is the risk that the chemicals used to clean the device will themselves pose a risk to patients by being absorbed, for example, by the plastic from which the device is made and then subsequently leaching out into a patient's body. There can also be the risk of deterioration of the device when exposed to cleaning or sterilization processes which could lead to device malfunction. When a manufacturer designs and develops a product, it determines the materials used and how the device is labeled.[8] In the U.S., to market a device as “reusable”, a manufacturer must provide increased data requirements and invest the resources necessary to demonstrate to FDA that the product can be safely reprocessed at the hospital level.[9] Unlike reusable devices, single-use devices are not sold with instructions on how they can be properly cleaned and sterilized nor have they been validated or tested for exposure to cleaning or sterilization processes, including chemical exposure and heat.

Commonly reprocessed SUDs

Commonly reprocessed medical devices include lower-risk, U.S. FDA Class I non-invasive devices such as sequential compression sleeves, tourniquet cuffs, and pulse oximeter sensors, to medium-risk, FDA Class II minimally invasive surgical devices including Ear, Nose and Throat microdebriders and cautery electrodes, laparoscopic graspers, scissors, forceps, scalpels, orthopedic blades, bits, burs, external fixation clamps, bolts and components, and electrophysiological cardiac catheters.[10] To date, the FDA has not approved for reprocessing any Class III, or higher risk, SUDs.

U.S. regulations for reprocessed SUDs

The Medical Device User Fee and Modernization Act of 2002 (MDUFMA), and Medical Device User Fee Stabilization Act of 2005 were signed into law on October 26, 2002, and August 1, 2005, respectively. All medical devices including reprocessed devices are subject to premarket review by the U.S. FDA, unless the agency has, by regulation, declared the device to be exempt from premarket requirements.[1] Unless exempt, the lower risk “Class I” and “Class II” devices, whether “original” or reprocessed, are required to have cleared premarket notification submissions (“510(k)s”).[11] With regard to premarket review, reprocessors are subject to more stringent regulation by FDA than are OEMs[12] because MDUFMA, require FDA to withdraw premarket notification exemptions for a significant number of previously exempt reprocessed devices, although “original” devices remain exempt from premarket review.[13] Reprocessors must also validate the cleaning and sterilization methods they seek to use in the reprocessing of a SUD, and include in their 510(k) submissions “validation data [...] regarding cleaning, sterilization, and functional performance” to show that the reprocessed device “will remain substantially equivalent [...] after the maximum number of times the device is reprocessed as intended”.[14] Both OEMs and reprocessors are subject to establishment registration and medical device listing; medical device reporting; medical device tracking; reports of corrections and removals, the quality system regulation (“QSR”); and labeling requirements.[6]

Safety and efficacy of reprocessed SUDs

FDA and the independent U.S. Government Accountability Office (GAO) have concluded that there is no evidence of harm to patients from FDA-regulated reprocessed SUDs. A 2008 GAO report found that of the over 320,000 adverse events filed with FDA between 2000 and 2006, only 65 adverse events “actually involved or were suspected to involve a reprocessed SUD and that the reprocessed SUD was one of several possible causal factors in the adverse event. In reviewing these 65 reports, FDA found that the types of adverse events reported to be associated with the use of reprocessed SUDs were the same types of events that were reported for new devices”.[15] The 2008 GAO report concluded: “After reviewing the available evidence – including FDA’s process for identifying and investigating device-related adverse events reported to involve reprocessed SUDs, peer-reviewed studies published since 2000, and the results of our and FDA’s consultations with hospital representatives – we found no reason to question FDA’s analysis indicating that no causative link has been established between reported injuries or deaths and reprocessed SUDs".[15] In a separate letter from FDA to Congressman Tom Davis and Harry Waxman dated January 23, 2006, FDA indicated that a total of 65,325 reports have been filed between 2003 and 2006 for the malfunction or injury associated with the first use of devices labeled for “single use.” The same search produced 176 cases of apparent malfunction or injury associated with reprocessed devices. Upon analysis of the latter reports, FDA determined that these adverse events were not related to the reprocessing of the “single use” device.[16]

Benefits of reprocessing

Economic

FDA-regulated reprocessed devices cost between 40%-60% of an original device.[17] Currently, reprocessors estimate that a typical 200-bed hospital, if taking advantage of a reprocessor’s full product line, can save between $600,000 and $1 million a year and divert between 5,000 and 15,000 pounds of waste from landfills.[18] According to a study by The Commonwealth Fund with funding from the Robert Wood Johnson Foundation and Health Care Without Harm in November 2012, they estimated that from hospitals implementing a reprocessing program, cost savings over five years was about $57 per procedure, and that if hospitals nationwide adopted an SUD reprocessing intervention, cost savings would be $540 million annually, or $2.7 billion over five years.[19]

Environmental

Regulated medical waste (RMW), or “red bag waste,” is a waste expenditure that typically costs hospitals 6 to 10 times more to dispose of than regular solid waste.[17] Among the inventory of devices reprocessed annually, ninety-five percent (95%) are recycled at the end of their life cycle rather than sent to landfills.[20] A variety of otherwise reprocessable raw materials that end up in a hospital’s RMW include stainless steel, aluminum, titanium, gold, polycarbonate and polyurethane. Reprocessing has allowed some hospitals to divert over 8,000 pounds of RMW from landfills each year, while larger systems can divert more than 50,000 pounds.[21]

Physician and Clinical Statements in Support

List of Known Regulated Third-Party SUD Reprocessing Vendors

Regulated Vendor Location
Hygia[30]Birmingham, AL
Medline ReNewal[31]Redmond, OR
NEScientific[32]Waterbury, CT
ReNu Medical[33]Everett, WA
SterilMed, Inc[34]Maple Grove, MN
Stryker Sustainability Solutions[35]Lakeland, FL
SureTek Medical[36]Greenville, SC
Vanguard (International)[37]Berlin, Germany

The Association of Medical Device Reprocessors (AMDR),[38] based in Washington, D.C., is the global trade association representing the interests of the third-party, “single-use” medical device reprocessing industry. AMDR was founded in 1997 and the third-party, single-use medical device reprocessors in the association now serve a majority of U.S. hospitals, including the country’s Honor Roll hospitals, as ranked by U.S. News & World Report.[39]

International Regulation of “Single-Use” Medical Device Reprocessing

The reprocessing of SUDs is commonplace worldwide. Even in developed nations, including those that have reprocessing prohibitions in place, hospitals routinely reuse SUDs in an unregulated manner. In many cases (particularly in Africa and Asia), uncontrolled reuse of such devices is relatively common, if not the norm.[40]

European Union

Currently, the European Union (EU) does not have a single policy regarding the reprocessing of SUDs. However, it is in the process of revising its Medical Device Directive. In 1993, at the issuance of the last Medical Device Directive, the issue of medical device reprocessing was identified as in need of additional clarification and the European Commission was instructed to submit a report on the issue by 2010.[41] In August, 2010, the Commission released its report,[42] highlighting the risks of unregulated reprocessing. Ultimately, the European Commission released its proposal[43] and the European Parliament approved its draft[44] (9 October 2013). Both legislative versions propose to regulate reprocessing as manufacturing. The European Council is now set to approve its own version of the legislation. Ultimately, both the Council and Parliament will have to agree, meeting in Trialogue with the European Commission, on the final version of the legislation. It is hoped the final agreement on the regulation will completed in 2015.

Until then, regulation of reprocessing activities is left to the individual Member States. Since 2001, Germany has had in place a regulatory framework that does not distinguish between the reprocessing of “reusable” and so-called “single-use” medical devices. The guidelines, therefore, allow for SUD reprocessing if conformance with certain standards is achieved. The German Medical Devices Law and the Medical Devices Operator Ordinance regulate the reprocessing of medical devices and in doing so refer to the mutual recommendation by the Robert Koch Institute (RKI) and the Federal Institute for Drugs and Medical Devices (BfArM) for the reprocessing of medical devices.[45] As a result, the RKI’s requirements must be observed.

Institutions, which want to reprocess single-use medical devices, must adopt and implement a quality management system according to DIN EN ISO 13485:2007. Compliance with the quality management requirements is monitored annually by “Notified Bodies” that have been accredited by the Central Authority of the Länder for Health Protection with regard to Medicinal Products and Medical Devices (ZLG).

Other Member States, such as the United Kingdom, Spain and France,[46] discourage or prohibit SUD reprocessing. The majority of Member States in Europe do not have any national regulations regarding reprocessing.[47]

Africa and the Middle East

The lack of resources, including medical devices and distribution channels, “necessitates the reuse of single-use devices” in much of Africa.[40] This includes the reuse of syringes and needles that have not been sterilized, and even rubber gloves. In the Middle East, available data indicates that reuse of SUDs is common throughout Arab countries (particularly for cardiac catheters), despite the absence of a regulatory framework. Reprocessing in both Africa and the Middle East is done at the user-facility level.[40]

Israel

Israel does not have regulations in place specific to the reprocessing of SUDs, but as a general matter, medical devices must be registered with the Ministry of Health (MOH) before they can be sold in the country. If a product is approved by the U.S. FDA, it will generally be registered by the MOH with no further testing requirements and, therefore, may be lawfully marketed in the country. Consistent with this policy, FDA-cleared reprocessed devices have been registered with MOH and are actively imported into the country.[48]

Asia/Japan

The reuse of SUDs in much of Asia is common, particularly for injection needles.[40] For the most part, there are no national regulations governing reuse of SUDs and, thus, third-party reprocessors do not offer their services in Asia. Rather, most reuse in Asia is conducted in an unregulated-manner at the user-facility level. Reprocessing is not currently regulated in Japan, but, available data indicates that the reuse of SUDs is relatively common. A 2003 survey found that 80 to 90 percent of hospitals reused SUDs.[40]

Australia

Australia enacted regulations regarding the reprocessing (“remanufacturing” in Australia) of SUDs in 2003.[49] Similar to the U.S., in Australia, all reprocessors (third-party, hospital, and OEM) must conform to medical device manufacturer requirements as regulated by the Therapeutic Goods Administration (TGA). Prior to implementation of these requirements, hospital reprocessing of SUDs was common.

Canada

Health Canada does not currently regulate reprocessed single-use devices or the third-party companies that are reprocessing these devices for Canadian hospitals. Whether or not to reprocess a single-use device is determined by the territorial and provincial health ministries as well as hospital boards.[50] A number of provinces have adopted similar positions that allow for the reprocessing of SUDs if the third party reprocessor is regulated by the FDA.

British Columbia

British Columbia issued a policy to its health authorities stating that by January 1, 2008, all health authorities must have eliminated the reprocessing and reuse of critical contact SUDs, unless they have been reprocessed by a licensed third-party reprocessor that is certified by a national regulatory authority such as Health Canada or the U.S. Food and Drug Administration.[50] The policy was revised in 2011 with additional information that sharps (e.g. scalpel blades, drill bits, saw blades, shavers) and needles shall be single-use and shall not be reprocessed.

Manitoba

Manitoba does not permit hospitals to reuse SUDs in-house, but does permit hospitals to contract with an FDA regulated vendor, among other requirements.[51]

Alberta

Alberta Health Services issued a policy in 2012 prohibiting the reuse and reprocessing of both critical and semi-critical single-use medical devices.[52]

Northwest Territories

Since 2005, the Northwest Territories have prohibited reprocessing. Specifically, the Northwest Territories Department of Health and Social Services revised its Hospital and Health Care Facility Standards Regulations to require that “a disposable device intended to be used on a patient during a single procedure shall not be used on a patient for more than one procedure and shall not be used on another patient." [50]

Ontario

In 2006, the Ontario Ministry of Health and Long Term Care endorsed a guidance document developed by its Provincial Infectious Diseases Advisory Committee (PIDAC) advising that critical and semi-critical SUDs must not be reprocessed and reused, unless the reprocessing is done by a licensed reprocessor.[50]

Saskatchewan

In 2013, The Saskatchewan Ministry of Health affirmed a policy outlining requirements for hospitals that reprocess SUDs. Consistent with the policies of other provinces, Saskatchewan requires, among other things, that hospitals outsource to an FDA regulated vendor.[53]

References

  1. 1 2 "Reprocessing of Single-Use Devices". Fda.gov. 2006-09-26. Retrieved 2013-10-24.
  2. "A Winning Approach | AMH Magazine - Preventing Infection in Medical Treatment". AMH Magazine. Retrieved 2013-10-24.
  3. Landro, Laura (2008-03-19). "Hospitals Reuse Medical Devices To Lower Costs - WSJ.com". Online.wsj.com. Retrieved 2013-10-24.
  4. Rodak, Sabrina (2013-06-14). "Reprocessing Saves Up to $20k Per OR Annually". Beckersasc.com. Retrieved 2013-10-24.
  5. 1 2 "Reuse of Medical Devices Labeled for Single-use". Fda.gov. Retrieved 2013-10-24.
  6. 1 2 3 "Enforcement Priorities for Single-Use Devices Reprocessed by Third Parties and Hospitals". Fda.gov. Retrieved 2013-10-24.
  7. "MDUFMA Frequently Asked Questions". Fda.gov. Retrieved 2013-10-24.
  8. "Reprocessed Single Use Devices: FDA Oversight Has Increased, and Available Information Does Not Indicate That Use Presents an Elevated Health Risk". GAO-08-147: Published: Jan 31, 2008. Publicly Released: Mar 3, 2008.
  9. "Draft Guidance for Industry and FDA Staff - Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling". Fda.gov. Retrieved 2013-10-24.
  10. "List of Single-Use Devices Known To Be Reprocessed or Considered for Reprocessing (Attachment 1)". Fda.gov. Retrieved 2013-10-24.
  11. "List 2 - Reprocessed Single-Use Devices Subject to Premarket Notification Requirements That Now Require the Submission of Validation". Fda.gov. Retrieved 2013-10-24.
  12. Testimony of Dr. Daniel Schultz, Director, CDRH, FDA (September 26, 2006) ("Congress mandated a number of new requirements for SUD reprocessors including, for certain SUDs, the pre-market submission of data to the agency that exceeded the requirements for the original manufacturers (OEMs)" http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/ReprocessingofSingle-UseDevices/ucm121067.htm
  13. "Guidance for Industry and FDA Staff - Medical Device User Fee and Modernization Act of 2002, Validation Data in Premarket Notification Submissions (510(k)s) for Reprocessed Single-Use Medical Devices". Fda.gov. Retrieved 2013-10-24.
  14. "68 Fed. Reg. 23139" (PDF). April 30, 2003. citing 21 U.S.C. § 360(o) (emphasis added). For a full description of the validation data reprocessors must submit on a premarket basis, including more particular guidance on cleaning, functional testing, and sterilization data requirements, see Guidance for Industry and FDA Staff: Medical Device User Fee and Modernization Act of 2002, Validation Data in Premarket Notification Submissions (510(k)s) for Reprocessed Medical Devices (Sept. 25, 2006), at 15
  15. 1 2 U.S. Government Accountability Office, GAO-08-147, Reprocessed Single-Use Medical Devices: FDA Oversight Has Increased, and Available Information Does Not Indicate That Use Presents an Elevated Health Risk (January 2008), at 1 http://www.gao.gov/new.items/d08147.pdf; See also, GAO’s 2000 report:
  16. "Letter from FDA to Congressman Tom Davis and Harry Waxman dated January 23, 2006" (PDF). Fda.gov. Retrieved 2013-10-24.
  17. 1 2 Sheehan, Kaeleigh. "Reprocess and Reuse". Greenhealth Magazine. Retrieved 2013-10-24.
  18. "Remanufactured Goods: An Overview" (PDF). Usitc.gov. Retrieved 2014-03-07.
  19. SuSan Kaplan, Blair Sadler, Kevin little, Calvin Franz, and Peter OrrisS "Can Sustainable Hospitals Help Bend the Health Care Cost Curve?". The Commonwealth Fund. Retrieved 2013-10-24.
  20. "AMDR Statement" (PDF). Usitc.gov. Retrieved 2014-03-07.
  21. "Healthier Hospitals Initiative 2012 Milestone Report" (PDF). Healthierhosptials.org. Retrieved 2014-03-07.
  22. "Letter from American College of Cardiology to Senator Richard Durbin in support of reprocessing" (PDF). 1999.
  23. "Reprocessing Advisory". ahrmm.org. Retrieved 2014-03-30.
  24. "Letter from American Hospital Association to Senator Thad Cochran in support of reprocessing" (PDF). 1999.
  25. "2000 Interim Meeting of the American Medical Association, Reports of the Council on Scientific Affairs" (PDF).
  26. "American Nursing Association 2010 House of Delegates Resolution: Safety and Effectiveness of Reprocessed Single Use Devices in Healthcare" (PDF). 2010.
  27. (PDF) http://amdr.org/documents/APICsingleusedevicepositionChangesIncluded-Final8_31_07-r.pdf. Missing or empty |title= (help)
  28. "Frequently Asked Questions | Practice Greenhealth". practicegreenhealth.org. Retrieved 2014-03-30.
  29. "Letter from Mayo Clinic to Secretary Tommy Thompson, Department of Health and Human Services" (PDF). 2001.
  30. http://www.hygia.net/ Official Website
  31. http://www.medlinerenewal.com/ Official Website
  32. http://www.mdreprocess.com/ Official Website
  33. http://renumedical.com/ Official Website
  34. http://www.sterilmed.com/ Official Website
  35. http://sustainability.stryker.com/ Official Website
  36. http://suretekmedical.com/ Official Website
  37. http://www.vanguard-healthcare.com/ Official Website
  38. http://www.amdr.org/ Official Website
  39. "Majority of U.S. News & World Report’s 2013–2014 "Honor Roll" Hospitals Support Reprocessing Industry" http://www.amdr.org/news/2013/08/reprocessing-industry-overwhelmingly-supported-by-u-s-news-world-report%E2%80%99s-2013-2014-%E2%80%9Chonor-roll%E2%80%9D-hospitals/ August 27, 2013
  40. 1 2 3 4 5 "What is the use? An international look at reuse of single-use medical devices". International Journal of Hygiene and Environmental Health. 213: 302–307. doi:10.1016/j.ijheh.2010.04.003.
  41. "Council Directive 93/42/EEC (14 June 1993), Article 12a, OJ L 169, 12.7.1993, p. 1."
  42. "Report from the Commission to the European Parliament and the Council: Report on the Issue of the Reprocessing of Medical Devices in the European Union, in Accordance with Article 12a of Directive 93/42/EEC, European Commission (27 August 2010)."
  43. "European Commission, Proposal for a REGULATION OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL, on medical devices, and amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 (26 September 2012)."
  44. "European Parliament, Report on the proposal for a regulation of the European Parliament and of the Council on medical devices, and amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 (COM(2012)0542 – C7-0318/2012 – 2012/0266(COD)), Committee on the Environment, Public Health and Food Safety (9 October 2013)."
  45. "Hygienic Requirements for Processing of Medical Devices: Recommendation by the Commission for Hospital Hygiene and Infection Prevention at the Robert Koch Institute (RKI) and the Federal German Institute for Medical Drugs and Medical Products (BfArM) Concerning the "Hygienic Requirements for Processing of Medical Devices," Robert Koch Institute: Recommendation (2001)."
  46. Commission Report, supra note 2, at 6. See also, Circulaire DGS/DH n 51 (December 29, 1994) Relative a l’utilisation des dispositifs medicaux steriles a usage unique dans les etblissements de santé publics et prives (France), Ministerio de sanidad y consume, Royal Decree 414.1996 Sec. 5, (July 9, 1999) (Spain) and MHRA Device Bulletin: DB 2006(04) Single-Use Medical Devices: Implications and Consequences of Reuse (UK)."
  47. Commission Report, supra note 2, at 6. See also, European Association of Medical Device Reprocessors."
  48. "Medical Device Regulatory Requirements for Israel,", U.S. International Trade Administration (May 2, 2005) http://www.ita.doc.gov/td/health/israelregs.html
  49. "Statement by the TGA on regulations for sterilisation of single use devices," Australian Government, Therapeutic Goods Administration (July 21, 2003) http://www.tga.gov.au/archive/media-2003-sud-030721.htm
  50. 1 2 3 4 CADTH Report "Issue Analysis Summary: The Reuse of Single-Use Medical Devices," Health Canada; Therapeutic Products Directorate (April 28, 2005) http://www.hc-sc.gc.ca/dhp-mps/alt_formats/hpfb-dgpsa/pdf/md-im/saprmd_ias_gcsrmm_raq_2005-06-09-eng.pdf. See also, "Reprocessing Single-Use Medical Devices: An Update of the Clinical Evidence and An Environmental Scan of Policies in Canada," Canadian Agency for Drugs and Technologies in Health (June 23, 2010) http://www.cadth.ca/media/pdf/ES8_Reprocessing_SUDs_e.pdf
  51. Manitoba Deputy Health Minister’s Office (March 21, 2013)
  52. Alberta Health Services. Single Use Medical Devices Policy. May 1, 2012 http://www.albertahealthservices.ca/6444.asp
  53. Saskatchewan Health, Deputy Minister’s Office (June 27, 2013)
This article is issued from Wikipedia - version of the 5/29/2016. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.