Computed tomography dose index

The computed tomography dose index (CTDI) is a commonly used radiation exposure index in X-ray computed tomography and is reported by the CT manufacturers to scan personnel for each exam. The CTDI can be used in conjunction with patient size to estimate the absorbed dose. The CTDI and absorbed dose may differ by more than a factor of two for small patients such as children.[1]

Definitions

It is defined by the US Department of Health and Human Services (HHS) as the average dose imparted by a single axial acquisition to a standard 100 mm pencil chamber dosimeter inside a PMMA phantom over the width of 14 CT slices:

where is the number of slices acquired, is the slice thickness and is the radiation dose measured at position along the scanner's main axis.

This measurement is most often made using a 100-mm standard pencil dose chamber as this is representative of a typical scan length:

.

The absorbed dose to water (used to refer back to patient dose) is typically measured in a cylindrical head (16 cm diameter) or body (32 cm diameter) phantom of length approximately 14–15 cm.

The dose distribution imparted by a CT scan is much more homogeneous than that imparted by radiography, but is still somewhat larger near the skin than in the centre of the body. The weighted CTDI was introduced to account for this:

using measurements acquired at central and peripheral positions in the head or body phantoms described above.

CTDI in helical CT

In helical CT, the pitch of the machine - a factor of the speed at which the couch travels through the gantry and the tube rotation frequency - also impacts on patient dose. The pitch factor, P, is defined as

where is the distance travelled by the couch during one full gantry rotation and is the beam collimation (single-slice CT) or the total thickness of all simultaneously acquired slices (multislice CT). The following quantity is therefore used to take account of pitch:

Similar measures with yet wider chambers are useful for CT systems with large numbers of detector rows.[2]

CTDI can also be measured with polymer gel dosimeters. [3]

Relation to DLP

The dose-length product (DLP) is a quantity defined for use in CT as

for and as described above ( is therefore the total scan length). This quantity is analogous to the dose-area product (DAP) used in planar radiography.

Notes

  1. McCollough, C. H.; Leng, S.; Yu, L.; Cody, D. D.; Boone, J. M.; McNitt-Gray, M. F. (18 April 2011). "CT Dose Index and Patient Dose: They Are Not the Same Thing". Radiology. 259 (2): 311–316. doi:10.1148/radiol.11101800. Retrieved 22 June 2012.
  2. Geleijns 2009
  3. Hill 2005